Skip to content

Public FDA record

Z-1193-2007: Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm).

Recall record Z-1193-2007 for Steris Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1193-2007
Company in FDA record
Steris Corp
FDA classification date
Aug 23, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard.

Source facts

Action
By telephone 3/22/2007, customers were instructed to cease using the product and a service technician would be dispatched immediately to repair the unit.
Initiated
2007-03-22
Root cause
Employee error
Terminated
2010-09-28
Distribution
Worldwide, including USA, Denmark, Belgium, Greece, India, and Russia.There are no Canadian accounts.
Recall number
Z-1193-2007
Recall status
Terminated
CDRH record ID
52810
FDA classified
2007-08-23
Recall event number
38031

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗