Skip to content

Public FDA record

Z-1193-2012: PointCare NOW CD4NOW Gold Reagent Pack, or Gold Pack, Part Number :30034 In Vitro Diagnostic : Assay re

Recall record Z-1193-2012 for Pointcare Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1193-2012
Company in FDA record
Pointcare Technologies, Inc.
FDA classification date
Mar 8, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Stability not assured at labeled temperature 2-42C

Source facts

Action
The firm, PointCare, issued a "Customer Service Advisory (CSA-00016)" dated November 2, 2011 to its customers to inform them of the new storage requirements of 2-30C as opposed to the previously claimed 2-42C. The letter describes the product, problem and action to be taken. The customers were instructed to correct the labeling on any Gold Reagent Packs and Gold Reagent and ensure that the Gold Reagent Packs and Gold Reagent they currently have in stock is stored within the 2-30C storage range. Future lots will be labeled with the updated storage temperatures. A member of our customer support team will provide additional information relative to this advisory. Should you have any questions or require further assistance, you may email us at customersupport@pointcare.net or call our offices directly at 1-508-281-6925.
Initiated
2011-11-02
Root cause
Storage
Terminated
2012-04-26
Distribution
International only: including countries of : Abuja, Haiti, Indonesia, Natal, Nigeria, Senegal, South Africa, Uganda, and United Kingdom.
Recall number
Z-1193-2012
Recall status
Terminated
CDRH record ID
107286
FDA classified
2012-03-08
Recall event number
61092

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗