Open original FDA record ↗Recall record
- FDA identifier
- Z-1193-2012
- Company in FDA record
- Pointcare Technologies, Inc.
- FDA classification date
- Mar 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Stability not assured at labeled temperature 2-42C
Source facts
- Action
- The firm, PointCare, issued a "Customer Service Advisory (CSA-00016)" dated November 2, 2011 to its customers to inform them of the new storage requirements of 2-30C as opposed to the previously claimed 2-42C. The letter describes the product, problem and action to be taken. The customers were instructed to correct the labeling on any Gold Reagent Packs and Gold Reagent and ensure that the Gold Reagent Packs and Gold Reagent they currently have in stock is stored within the 2-30C storage range. Future lots will be labeled with the updated storage temperatures. A member of our customer support team will provide additional information relative to this advisory. Should you have any questions or require further assistance, you may email us at customersupport@pointcare.net or call our offices directly at 1-508-281-6925.
- Initiated
- 2011-11-02
- Root cause
- Storage
- Terminated
- 2012-04-26
- Distribution
- International only: including countries of : Abuja, Haiti, Indonesia, Natal, Nigeria, Senegal, South Africa, Uganda, and United Kingdom.
- Recall number
- Z-1193-2012
- Recall status
- Terminated
- CDRH record ID
- 107286
- FDA classified
- 2012-03-08
- Recall event number
- 61092