Open original FDA record ↗Recall record
- FDA identifier
- Z-1193-2014
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Mar 12, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Ortho Clinical Diagnostics has determined that the VITROS Immunodiagnostic Products 25-OH Vitamin D Total Calibrators, Lots 0091 through 0235, used with VITROS Immunodiagnostic Products 25-OH Vitamin D Total Reagent Packs, may generate negatively biased results. As a result, users must immediately discontinue using all current lots of VITROS 25-OH Vitamin D Total Calibrators and Reagent Packs.
Source facts
- Action
- On 1/22/2014, Customer Letters were sent via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to US consignees to inform them of the issue and advise them to discontinue use of the affected lot(s) and to inform their Laboratory Medical Director if they suspect that a previously reported result may have been affected. On 1/22/2014, Distributor Letters were sent via US Postal Service Priority Mail and/or FedEx overnight courier to US distributor consignees and instructed them to notify their consignees of the issue and actions.
- Initiated
- 2014-01-15
- Root cause
- Component design/selection
- Terminated
- 2017-04-03
- Distribution
- Nationwide
- Recall number
- Z-1193-2014
- Recall status
- Terminated
- CDRH record ID
- 125454
- FDA classified
- 2014-03-12
- Recall event number
- 67439