Z-1193-2019: EV1000 Clinical Platform, Models: EV1000A, EV1000NI, EV1000CS Product Usage: EV1000A Monitor is used to

Recall record Z-1193-2019 for Edwards Lifesciences, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1193-2019
Company in FDA record
Edwards Lifesciences, LLC
FDA classification date
May 21, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for an electrical short circuit leading to possible fire hazard due to liquid ingress into AC power outlets.

Source facts

Action
On 03/21/19, Urgent Medical Device Safety Notification - Action Required notices were mailed to consignees. The firm strongly advised that consignees be cautious that IV bags and other fluids need to be positioned away from AC power connectors in order to avoid liquid ingress. In addition, consignees were asked to follow installation instructions in the operator's manual and to complete and return the response form. On 03/30/19, Urgent Medical Device Safety Notification - Action Required Addendum was mailed informing customers that if they require assistance they may call Technical Support is 800-822-9837. On 05/10/19, Urgent Medical Device Safety Notification - Action Required Addendum was mailed informing customers that improper mounting of the power supply could result in liquid ingress into the adaptor causing a short. Additional potential causes, are as follows: Reports of Burnt Cables: - We have reports of cables that appear burnt, without smoke, sparks or fire. We continue to investigate to identify all potential causes. One potential cause is the presence of conductive material in the AC inlet of the power supply. - Please assure that there is no conductive material in or near the AC inlet when connecting the AC power cord to the device power supply, and that both the AC power cord and inlet are clean and dry. Reports of Smoke, Sparks or Fire due to Cleaning or Decontamination/Disinfection: - The product instructions indicate that product should be kept dry and the cables should not be immersed in any liquid solution or liquids allowed to penetrate connectors. - Another opportunity for liquid ingress into the power supply could be the cleaning and disinfection process. Although the cleaning instructions in the Operators Manual indicate both to only wipe with a damp cloth and to dry the cable after cleaning, user reports indicate that this process is not always followed. - Please assure that the cleaning instructions as noted in the Oper
Initiated
2019-03-21
Root cause
Use error
Terminated
2023-11-28
Distribution
Worldwide Distribution - US Nationwide : FL, MD, NY, TX, VA, CO, OK, MO, CA, DE, TN, NC, DC, PA, MA, OR, SC, OH, WI, KS, NV, MI, IL, KY, GA, NJ, WA, CT, IN, MN, WV, LA, MS, AL, SD, NE, IA, AZ, ID, HI, UT, ME,MT, NM, AK, NH, RI, AR, ND, WY, PR. Foreign (OUS): IT, FI, CO, KR, JP, PL, BE, GR, NL, IL, AT, DK,CA, GB, BR, SA, SE, DE, AU, FR, MX, IN, IC, ZA, ES, TR, NZ, MY, IE, SG, CZ, SI, PT, CH, RE, LK, CR, AR, HK, NO, PY, PA, MT, EG, NP, TN, RU, RO, NA,CL, BH, IR, PH, TH, KZ, DZ, BG, EE, KW, VN, OM, PE, EC, AE, MA, VE, BO, QA, DO, GT, BY, JO, LT, SK, HU, CY, MN, TW, CN
Recall number
Z-1193-2019
Recall status
Terminated
CDRH record ID
171641
FDA classified
2019-05-21
Recall event number
82456

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗