Open original FDA record ↗Recall record
- FDA identifier
- Z-1194-2016
- Company in FDA record
- QiG Group LLC
- FDA classification date
- Mar 18, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.
Source facts
- Action
- Consignee was sent on 1/26/2016 a Algostim, LLC "Urgent Field safety Notice" dated January 21, 2016. The letter described the problem and the product involved in the recall. The letter provided Background information as well as described the Clinical Implications, Rate of Occurrence, Recommendations and Mitigations. Advised consignee to deliver the respective letter to each physician who have implanted affected product and collect signed Response forms.
- Initiated
- 2016-01-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2017-04-05
- Distribution
- Distributed in Germany and Luxembourg.
- Recall number
- Z-1194-2016
- Recall status
- Terminated
- CDRH record ID
- 143234
- FDA classified
- 2016-03-18
- Recall event number
- 73150