Open original FDA record ↗Recall record
- FDA identifier
- Z-1194-2019
- Company in FDA record
- Roche Diagnostics Corporation
- FDA record creation date
- Apr 18, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The reagent lot is exhibiting calibration failures, QC imprecision, and elevated QC recovery due to elevated pH level causing higher absorbance values.
Source facts
- Action
- Urgent Medical Device Correction notification letters dated 3/18/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: Stop using and discard all Albumin BCP reagent cassettes from lot number 35651401 (catalog number 05975573190) remaining in your inventory according to your local guidelines. Actions Required " Complete all sections of the enclosed fax form and fax it to 1-833-254-2600 or email it to Roche10060@stericycle.com even if you are not requesting product replacement. " Discard remaining affected product in your inventory according to local guidelines. " File this Urgent Medical Device Correction (UMDC) for future reference.
- Initiated
- 2019-03-18
- Root cause
- Process control
- Terminated
- 2020-08-28
- Distribution
- US Nationwide distribution in the states of: GA, IL, KY, MI, and TX.
- Recall number
- Z-1194-2019
- Recall status
- Terminated
- CDRH record ID
- 171518
- FDA record created
- 2019-04-18
- Recall event number
- 82388