Skip to content

Public FDA record

Z-1194-2019: Albumin BCP Product Usage: ALBP is an in vitro test used for the quantitative determination of albumin

Recall record Z-1194-2019 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1194-2019
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Apr 18, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reagent lot is exhibiting calibration failures, QC imprecision, and elevated QC recovery due to elevated pH level causing higher absorbance values.

Source facts

Action
Urgent Medical Device Correction notification letters dated 3/18/19 were sent to customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to: Stop using and discard all Albumin BCP reagent cassettes from lot number 35651401 (catalog number 05975573190) remaining in your inventory according to your local guidelines. Actions Required " Complete all sections of the enclosed fax form and fax it to 1-833-254-2600 or email it to Roche10060@stericycle.com even if you are not requesting product replacement. " Discard remaining affected product in your inventory according to local guidelines. " File this Urgent Medical Device Correction (UMDC) for future reference.
Initiated
2019-03-18
Root cause
Process control
Terminated
2020-08-28
Distribution
US Nationwide distribution in the states of: GA, IL, KY, MI, and TX.
Recall number
Z-1194-2019
Recall status
Terminated
CDRH record ID
171518
FDA record created
2019-04-18
Recall event number
82388

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗