Patient positioning errors may occur because of mechanical variations during product installation.
Source facts
Action
The firm issued a notification letter to all customers April 18, 2007. The firm intends to arrange for service personnel to contact affected sites to verify the calibration of the OBI system, and make arrangements to correct any issues. The firm will also update installation manuals to clearly describe the OBI installation and calibration requirements.
Initiated
2007-04-18
Root cause
Device Design
Terminated
2010-08-12
Distribution
Worldwide, including USA, Puerto Rico, Australia, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Ireland, Italy, Japan, Luxembourg, New Zealand, Norway, Poland, Portugal, Spain, Sweden, Switzerland, Netherlands, and United Kingdom.
Recall number
Z-1195-2007
Recall status
Terminated
CDRH record ID
53471
FDA classified
2007-08-29
Recall event number
38349