Mis-labeing; The instructions for use include a use for which the device is not intended. (Device not approved to capture a bone biopsy)
Source facts
Action
Consignees were notified via Medical Device Recall Notification letter dated 2/13/08 to read the enclosed instructions for use, remove all old instructions for use at the consignee location and replace them with the new instructions for use, and to discard the old instructions for use.
Initiated
2008-02-13
Root cause
Labeling Change Control
Terminated
2008-12-30
Distribution
Nationwide.
Recall number
Z-1195-2008
Recall status
Terminated
CDRH record ID
68287
FDA classified
2008-03-11
Recall event number
46726