Open original FDA record ↗Recall record
- FDA identifier
- Z-1195-2009
- Company in FDA record
- GE Medical Systems, LLC
- FDA classification date
- May 19, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
GE Healthcare has become aware of a potential image storage failure on Innova 2121 IQ/ 3131 IQ cardiovascular imaging systems that may impact patient safety. This intermittent communication failure on the image storage chain may result in the following situations: Record abort, Record not available, Fluorostore function not available, Review not available, Boot/Reset failure. Fluoroscopy imag
Source facts
- Action
- A GE Healthcare "Urgent Medical Device Correction" letter dated February 2, 2009 was sent to consignees via mail. The letter was directed to Hospital Administrator/Risk Manger, Managers of Radiology/Cardiology and Radiologists/Cardiologists addressing the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. Further questions should be addressed to GE Healthcare at 1-800-437-1171.
- Initiated
- 2008-11-21
- Root cause
- Device Design
- Terminated
- 2013-11-21
- Distribution
- Worldwide Distribution -- USA including states of CA, CO, FL, HI, IA, IL, IN, KS, MD, MA, MN, MO, NY, OH, OR, PA, TN, TX, WA, WV, WI and VA and countries of the United Kingdom, Romania, Japan, Taiwan, Bahrain, France, Malaysia, India, Canada, Chile, Indonesia, Ireland and Italy.
- Recall number
- Z-1195-2009
- Recall status
- Terminated
- CDRH record ID
- 79028
- FDA classified
- 2009-05-19
- Recall event number
- 51193