- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There exists the possibility that the ADU Fresh Gas Control unit (A-FGC1) can shut off during normal handling of the anesthesia machine and result in no anesthetic agent output.
Source facts
- Action
- An Urgent Medical Device Correction letter dated March 21, 2006, was sent to customers. It warns of the potential for no anesthetic agent output. The letter contains short term recommendations for use of the device until a GE Healthcare representative contacts the facility to schedule a vist and implelment corrective action.
- Initiated
- 2006-03-22
- Root cause
- Other
- Terminated
- 2007-05-29
- Distribution
- World wide-(including the States of IA, CA, MI, FL, MI, TX, IL, WI, OH, and countries of Finland, Singapore, Austria, South Africa, Sweden, Israel, Norway, Taiwan, Germany, Turkey, Portugal, Italy, Switzerland, Great Britain, and France
- Recall number
- Z-1196-06
- Recall status
- Terminated
- CDRH record ID
- 44224
- FDA classified
- 2006-07-12
- Recall event number
- 34631