Open original FDA record ↗Recall record
- FDA identifier
- Z-1196-2012
- Company in FDA record
- Philips Medical Systems Nederlands
- FDA classification date
- Mar 9, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Displayed skin dose rate is calculated using an incorrect point of reference, resulting in too low displayed values.
Source facts
- Action
- Philips Healthcare issued Urgent Field Safety Notices dated February 21, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. No action is necessary by the customer/user. A mandatory Field Safety Correction (FCO) Action will be issued to solve the problem. The FCO will consists of a Software update. For further support customers were instructed to contact their Phillips representative. For questions regarding this recall call 978-687-1501.
- Initiated
- 2012-02-24
- Root cause
- Software design
- Terminated
- 2013-03-26
- Distribution
- Worldwide Distribution - USA (nationwide) and the following countries: Canada, Australia Austria Belgium Bahrain Czech Republic Denmark Estonia France Finland Germany Hong Kong Israel Italy Latvia Luxemburg Malaysia Mexico Netherlands New Zealand Norway Russian Fed SAUDI ARABIA Serbia Slovakia Solvenia South Africa Spain Switzerland Sweden Syria Thailand Turkey United Kingdom
- Recall number
- Z-1196-2012
- Recall status
- Terminated
- CDRH record ID
- 107690
- FDA classified
- 2012-03-09
- Recall event number
- 61267