Open original FDA record ↗Recall record
- FDA identifier
- Z-1196-2015
- Company in FDA record
- Navilyst Medical, Inc
- FDA classification date
- Feb 26, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.
Source facts
- Action
- Navilyst Medical sent an Urgent Medical Device Recall Notification letter dated January 9, 2015 to US customers (end users) via Federal Express. On January 20, 2015, Recall Notification Packages were sent via Federal Express to 2 International consignees (distributors). the letter identified the affected product, problem and actions to be taken. Customers are instructed to: " Immediately remove the recall product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). " Segregate this product in a secure location for return to Navilyst Medical. " Immediately forward a copy of this recall notification to all sites to which you have distributed affected product. " Complete and return the Reply Verification Tracking Form.
- Initiated
- 2015-01-09
- Root cause
- Packaging
- Terminated
- 2017-03-10
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Canada, and Guam.
- Recall number
- Z-1196-2015
- Recall status
- Terminated
- CDRH record ID
- 133756
- FDA classified
- 2015-02-26
- Recall event number
- 70322