Z-1197-2024: Ref No: INS7250 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventric

Recall record Z-1197-2024 for Integra LifeSciences Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1197-2024
Company in FDA record
Integra LifeSciences Corp.
FDA classification date
Feb 28, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.

Source facts

Action
Integra issued 1st Notification- Urgent: Voluntary Medical Device Recall letter on 01/19/24 via fedex overnight. Letter states reason for recall, health risk and action to take: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service and quarantine. 2. If you do have affected product, check the box on the enclosed Acknowledgement Form I do have affected product. Record the total quantity of the affected product that you have. 3. Forward this notice to those who utilize the device so they are aware of this recall and can identify any affected devices that may remain in clinical areas. 4. If you do not have affected product, check the box, I do not have affected product. 5. Complete the rest of the Acknowledgement Form and return to FCA1@integralife.com or FAX to 1-609-750-4220. 6. Keep a copy of the form for your records. 7. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. You can request a credit or a replacement for the quantities returned (based on inventory availability). Actions to be Taken by Distributors/Sales Reps: 1. If you do have the affected product listed in Table 1, remove the product from further distribution. 2. Complete the attached Acknowledgment Form. 3. If you do have affected product, check the box I do have affected product. Record the total quantity of affected product you have. 4. If you do not have affected product, check the box, I do not have affected product. 5. After completion, please email the Acknowledgement Form to FCA1@integralife.com or FAX to 1-609-750-4220. 6. Keep a copy of the form for your records. 7. Please check your customer traceability records for shipments of above catalog and lot numbers. 8. Forward a copy of the enclosed Field Safety Notice to any of your customers that have purchased the above catalog
Initiated
2024-01-19
Root cause
Process control
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Bahamas, Belgium, Canada, Chile, France, Germany, Greece, Ireland, Israel, Poland, Portugal, Slovenia, South Africa, Spain, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Recall number
Z-1197-2024
Recall status
Open, Classified
CDRH record ID
205787
FDA classified
2024-02-28
Recall event number
93881

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗