Unexpected powering down/failure to boot up: A problem with a transistor in the heater/temperature sensor component in the device may cause the device to power off during surgery or to fail to boot up at all. The risk includes delay and interruption of ongoing surgical procedures.
Source facts
Action
Recall initiated 12/29/2009. Sales representatives were notified via phone and e-mail, while product advisory notices were sent to consignees via FedEx. Consignees are provided with a response form, and are instructed to run a software correction.
Initiated
2008-12-29
Root cause
Component design/selection
Terminated
2011-03-24
Distribution
Product was distributed throughout the US and to Argentina, Brazil, Canada, Chile and Latin America, as well as to 16 government facilities.
Recall number
Z-1198-2010
Recall status
Terminated
CDRH record ID
77163
FDA classified
2010-03-23
Recall event number
50777