Skip to content

Public FDA record

Z-1198-2023: Biofinity XR Toric Single Diagnostic lens- Soft (hydrophilic) Contact Lens (extended wear)

Recall record Z-1198-2023 for CooperVision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1198-2023
Company in FDA record
CooperVision, Inc.
FDA classification date
Mar 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Manufactured with misaligned axis resulting in lenses with the incorrect power, users may experience poor visual acuity

Source facts

Action
Coopervision issued Urgent Medical Device Correction letter on 1/26/23 to direct consignees requesting end users be notified. Letter states reason for recall, health risk and action to take: Please examine your inventory/product shelf for any of the provided affected lot numbers. The lot number can be found on the back panel of the carton below the expiration date. Below is an example and where to locate the lot number. " Stop any further distribution of product from the designated lot numbers; " We recommend you contact your Eye Care Practitioners or patients who have received the affected product. Please instruct Eye Care Practitioners or patients to return any affected product to you for replacement or credit. " Enclosed you will find a template that can be customized by the ECP for distribution to their patients. Please note that if a credit is due to you, this will automatically be applied to your account once we have received and validated your Correction Response Form . This action is being reported to applicable health authorities and CooperVision may be required to identify any consignees that do not respond. Should you wish to discuss this action, please contact your business development manager or our customer services team at: 800-341-2020, 9:00 AM 7:00 PM ET, Monday Friday.
Initiated
2023-01-26
Root cause
Employee error
Distribution
US Nationwide distribution in the states of CA, OH, NC, PA.
Recall number
Z-1198-2023
Recall status
Open, Classified
CDRH record ID
198327
FDA classified
2023-03-01
Recall event number
91607

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗