Skip to content

Public FDA record

Z-1199-06: Ossur Total Knee 2000, Model Number TK 2000, external prosthetic knee.

Recall record Z-1199-06 for Ossur North America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1199-06
Company in FDA record
Ossur North America Inc
FDA classification date
Jul 4, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The company's initiation of the recall is based on a finding that some units of the Total Knee device may contain faulty pins based in the axis of the knee. Small cracks propagate through the pin cross section until the pin breaks, severely compromising the knee function.

Source facts

Action
On 3/29/06 a press release was posted on OandP.com, an industry specific web site. An announcement listing involved serial numbers and contact phone numbers was also posted on the company''s web site, ossur.com. A two stage mailing has been constructed. The first letter sent to all customers March 30, 2006, describes the problem and models affected. The second letter, March 31, 2006, is customer specific and lists all of the involved serial numbers by model which were purchased by that customer. Telephone calls will be placed to customers who do not respond to the mailings.
Initiated
2006-03-29
Root cause
Other
Terminated
2008-08-08
Distribution
Distribution worldwide in Austria, Canada, Columbia, Hong Kong, Japan, Korea, Kuwait, Malaysia, Mexico, Taiwan, & New Zealand. Involved states are: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, AND WY.
Recall number
Z-1199-06
Recall status
Terminated
CDRH record ID
45390
FDA classified
2006-07-04
Recall event number
35132

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗