Open original FDA record ↗Recall record
- FDA identifier
- Z-1199-2017
- Company in FDA record
- Acumed LLC
- FDA classification date
- Feb 14, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.
Source facts
- Action
- Acumed sent an Urgent Medical Device Recall letter dated January 16, 2017, to all affected customers on January 19, 2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or via phone 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada) have to obtain a Return Material Authorization (RMA) number.
- Initiated
- 2017-01-19
- Root cause
- Mixed-up of materials/components
- Terminated
- 2017-05-08
- Distribution
- Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain
- Recall number
- Z-1199-2017
- Recall status
- Terminated
- CDRH record ID
- 152823
- FDA classified
- 2017-02-14
- Recall event number
- 76332