Open original FDA record ↗Recall record
- FDA identifier
- Z-1200-2018
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Mar 26, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The ECG signal may flatline due to electromagnetic interference during the use of electrosurgical equipment.
Source facts
- Action
- Undated recall letters were issued via email (if they had an email address for the customer) or via certified mail (if no email address was available) on 3/5/2018.
- Initiated
- 2018-03-05
- Root cause
- Device Design
- Terminated
- 2021-02-26
- Distribution
- Distribution was made to medical facilities in AZ, CA, CT, FL, IL, IN, KS, KY, MI, MN, MO, NJ, NY, NC, OH, PA, TX, VA, and WA. There was no military/government distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Bosnia Herzeg, Colombia, Croatia, Denmark, France, Germany, India, Ireland, Israel, Italy, Japan, Mexico, New Zealand, Philippines, Poland, Republic Korea, Romania, Singapore, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, and Vietnam.
- Recall number
- Z-1200-2018
- Recall status
- Terminated
- CDRH record ID
- 162658
- FDA classified
- 2018-03-26
- Recall event number
- 79568