Reason stated in the source
There exists a potential situation with the crimp connection at the pump head assembly. The potential exists that the reverse osmosis pump head wiring could build resistance resulting in resistive heating of the surrounding components. If this occurs, the pump head assembly may experience melting of the plastic components and potentially create smoke/fire and shut down of the unit.
Source facts
- Action
- Mar Cor Technicians started visiting the consignees to make the correction, in June 2012 Mar Cor called the consignee prior to the visit. An Urgent Medical Device correction letter was sent to customers beginning 2/27/2014. the letter identified the affected units, described the issue, stated that a correction has been, or will be made to the devices and asked for a response form be sent back to the firm.
- Initiated
- 2012-06-27
- Root cause
- Component design/selection
- Terminated
- 2014-08-22
- Distribution
- nationwide. No distribution outside the US.
- Recall number
- Z-1201-2014
- Recall status
- Terminated
- CDRH record ID
- 125653
- FDA classified
- 2014-03-13
- Recall event number
- 67516