Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.
Source facts
Action
Terumo Cardiovascular Systems has contact all affected parties (user facilities and consignees) via a Customer Notification Letter that advises of the recall activity.
Initiated
2014-08-20
Root cause
Device Design
Terminated
2017-02-15
Distribution
Terumo Cardiovascular has identified 16 domestic consignees and 30 international consignees .
Recall number
Z-1201-2017
Recall status
Terminated
CDRH record ID
152881
FDA classified
2017-02-14
Recall event number
76344
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.