Z-1201-2023: MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laborato

Recall record Z-1201-2023 for Biomerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1201-2023
Company in FDA record
Biomerieux Inc
FDA classification date
Mar 2, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.

Source facts

Action
BioMerieux, US subsidiary issued an URGENT FIELD SAFETY NOTICE to its consignees beginning on 02/06/2023 by email and letter. The notice explained the issue with the product and potential risk, and requested the use take the following actions: "1. Confirm this letter has been distributed to and reviewed by all appropriate personnel within your organization. 2. If you are using VITEK MS as part of your laboratory workflow AND you currently use MYLA AST Filter rules disable current MYLA AST Filter rules. Review AST results to determine if some results should be suppressed from reporting. 3. If you are currently using MYLA AST Filter rules WITHOUT VITEK MS as part of your laboratory workflow, there is no impact and no action is required. 4. If you are not currently using MYLA AST Filter rules WITH or WITHOUT VITEK MS as part of your laboratory workflow, there is no impact and no action is required. 5. If you PLAN to use MYLA AST Filter rules WITH VITEK MS as part of your laboratory workflow in the future, please wait to begin usage of the MYLA AST Filter rules until MYLA version 4.9.1 has been installed on your systems. MYLA AST Filter rules may be activated at any time for workflows that do not include VITEK MS. 6. Once the patch to MYLA version 4.9.1 has been installed, usage of MYLA AST Filter Rules with VITEK MS as part of the laboratory workflow may resume. 7. Complete the attached Acknowledgement Form and return it to your local bioMrieux representative as soon as possible."
Initiated
2023-02-06
Root cause
Software design
Distribution
US Nationwide distribution to the states of Alabama, California, North Carolina, and Puerto Rico.
Recall number
Z-1201-2023
Recall status
Open, Classified
CDRH record ID
198358
FDA classified
2023-03-02
Recall event number
91617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗