Open original FDA record ↗Recall record
- FDA identifier
- Z-1202-06
- Company in FDA record
- Spacelabs Medical Incorporated
- FDA classification date
- Jul 6, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.
Source facts
- Action
- On May 5, 2006 a letter was issued to all consignees, via return receipt requested. The letter advised consignees of the issue and provided some ways to mitigate the risk. Anyone experiencing the anomaly is to document the failure and report it to the firm. A field service representative will visit each consignee to optimize the antenna system to reduce intermodulations. After determining the root cause and validating a fix, the firm issued a second letter dated August 21, 2006 to their consignees. The letter advised consignees of three issues that needed 'improvement' and stated a service engieneer will schedule a time to visit and install the improvements.
- Initiated
- 2006-05-05
- Root cause
- Other
- Terminated
- 2010-12-27
- Distribution
- The systems were distributed to 22 hospitals/medical centers nationwide and to 2 Spacelabs personnel. There is no international distribution.
- Recall number
- Z-1202-06
- Recall status
- Terminated
- CDRH record ID
- 45982
- FDA classified
- 2006-07-06
- Recall event number
- 35391