Z-1202-06: Spacelabs Medical 1400 MHz Telemetry System,. Model 91341-09. Telemetry System consists of Model 91341-

Recall record Z-1202-06 for Spacelabs Medical Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1202-06
Company in FDA record
Spacelabs Medical Incorporated
FDA classification date
Jul 6, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for patient waveforms to move to an open receiver module or to an occupied receiver causing the intermittent inappropriate monitoring of both patients for several minutes.

Source facts

Action
On May 5, 2006 a letter was issued to all consignees, via return receipt requested. The letter advised consignees of the issue and provided some ways to mitigate the risk. Anyone experiencing the anomaly is to document the failure and report it to the firm. A field service representative will visit each consignee to optimize the antenna system to reduce intermodulations. After determining the root cause and validating a fix, the firm issued a second letter dated August 21, 2006 to their consignees. The letter advised consignees of three issues that needed 'improvement' and stated a service engieneer will schedule a time to visit and install the improvements.
Initiated
2006-05-05
Root cause
Other
Terminated
2010-12-27
Distribution
The systems were distributed to 22 hospitals/medical centers nationwide and to 2 Spacelabs personnel. There is no international distribution.
Recall number
Z-1202-06
Recall status
Terminated
CDRH record ID
45982
FDA classified
2006-07-06
Recall event number
35391

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗