Z-1202-2021: Cascadion SM Clinical Analyzer, REF 99990000 SN Type: 9910

Recall record Z-1202-2021 for Thermo Fisher Scientific. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1202-2021
Company in FDA record
Thermo Fisher Scientific
FDA record creation date
Mar 10, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to software defect, under certain assay parameters, false Vitamin D results may be reported. The system reports a false result by not quantitating the correct analyte peak in sample chromatogram. The resulting sample would show a Vitamin D-concentration that is an unusual situation for human serum and plasma samples and would indicate a severe Vitamin D deficiency.

Source facts

Action
On 02/01/2021, the firm sent an "URGENT - MEDICAL DEVICE RECALL" Notification Letter to inform customers that the Recalling Firm has identified that there is a software anomaly in the Analyzer Software subcomponent called Suhaili Service (version 2.0.3 and earlier) that is affecting the Analyzer. Suhaili Service controls raw data processing based on assay configuration data from Sample Prep software. Processed results are sent to Sample Prep software for patient result reporting. This anomaly, as a result of this interface, has a potential risk of reporting wrong patient results for Vitamin D. Customers are informed, As an immediate correction the following instructions must be followed with Cascadion SM Clinical Analyzer Software versions 2.0.2 and earlier until the installation of the new software version 2.1. 1. Please distribute this information immediately to any staff that may be impacted by this issue. 2. Please change the Quick Connect Cartridge Cs on the instrument to new ones. 3. After the cartridge change, if a sample is reported to have concentration of total 25-hydroxy Vitamin D below assay measuring range, follow these instructions to prevent the error from being reported: a. Repeat the sample analysis with a Cascadion SM Clinical Analyzer b. In case the repeated result is below assay measuring range, the sample should be rerun on another assay and platform. 4. Retain a copy of this letter for your laboratory records. 5. As appropriate, contact your Medical Professional for evaluation of further action. 6. Please, fill out the MEDICAL DEVICE RECALL- Response Form and return it within 5 days of the date of this letter to your manufacturer as instructed in the form and as listed: Email: vigilance.clinical.fi@thermofisher.com The actions to be taken by the Recalling Firm: 1. The Recalling Firm has informed the appropriate Regulatory Agencies. 2. When, available, the final corrective action will be to update Clinical Analyzer software v2.1
Initiated
2021-02-01
Root cause
Software design
Terminated
2024-01-10
Distribution
U.S.: Texas O.U.S.: Belgium, France, Germany, Italy, Spain, Switzerland and United Kingdom
Recall number
Z-1202-2021
Recall status
Terminated
CDRH record ID
185792
FDA record created
2021-03-10
Recall event number
87309

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗