Skip to content

Public FDA record

Z-1202-2026: Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S. 3200TS Tourniquet Systems Model/Catalog Number

Recall record Z-1202-2026 for Zimmer Surgical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1202-2026
Company in FDA record
Zimmer Surgical Inc
FDA classification date
Jan 28, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.

Source facts

Action
On December 30, 2025, the firm began notifying consignees through Urgent Medical Device Recall letters. The firm is conducting this recall in two phases. In Phase 1, which began on 12/30/2025, consignees are informed that the screen may freeze or stop responding to touch when used in certain international languages. Users may continue to use the device by utilizing the English language mode. Use of the device in an affected language is not advised, and may cause the issue to occur. ATS5000 Affected Languages: Czech, Danish, Finnish, French, Italian, Lithuanian, Norwegian, Romanian, Slovak, Slovenian, and Turkish Unaffected Languages: English, Chinese (simplified), Korean, Japanese, Greek, Bulgarian, and Russian All devices manufactured between April 2024 - September 2025 with software version v2.04 are within scope of the recall. Affected devices will require an upgrade to the new software version v2.05 when it is available in early 2026. The firm will issue a removal notice (Phase 2) once the new software version is available. Customers are instructed to not return affected devices before the initiation of Phase 2.
Initiated
2025-12-24
Root cause
Under Investigation by firm
Distribution
International distribution to the countries of Canada and EMEA only.
Recall number
Z-1202-2026
Recall status
Open, Classified
CDRH record ID
217977
FDA classified
2026-01-28
Recall event number
98235

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗