- FDA identifier
- Z-1203-2007
- Company in FDA record
- MAQUET Inc.
- FDA classification date
- Sep 7, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Overcharging of batteries and the release of acid fumes
Source facts
- Action
- Safety Alert letters were sent out by FedEx tracking. A Service Express letter will be sent out emphasizing that only qualified batteries must be used in KION equipment and will list all qualified battery types. A Field Action - update of existing operating manuals will include warning on battery charging on KION. This action is currently scheduled to start 9/1/07 and be completed in US by 10/1/07. After FDA consultation the recalling firm issued a 9/5/2007 URGENT DEVICE CORRECTION Letter stating that KION will visit consignees and replace batteries older than three years, verify that the batteries are ones of the approved type, and verify battery condition.
- Initiated
- 2007-07-10
- Root cause
- Other
- Terminated
- 2007-12-07
- Distribution
- Nationwide: 2 units at the National Institute of Health in Bethesda, MD and 7 units at Oconee Regional Medical Center, Milledgeville, GA
- Recall number
- Z-1203-2007
- Recall status
- Terminated
- CDRH record ID
- 53647
- FDA classified
- 2007-09-07
- Recall event number
- 38428