Reason stated in the source

Small black plastic particle was observed in the left ventricle during implant procedure.

Source facts

Action
On 03/01/19, sales representatives started visiting and telephoning consignees informing them to quarantine affected devices. The firm requested that consignees complete and return the Device Retrieval Template. On 03/08/19, Urgent Medical Device Recall notices were hand delivered to all impacted accounts. Customers were advised to stop use, quarantine, and return any units on hand. For patients, who's implant procedure included the use of affected devices, firm recommends close vigilance for potential harm, which could include concerns such as infection or thromboembolism. The coring knife is not impacted. Firm requests that customers complete and return the acknowledgement form. Firm recommends that customer with additional questions call 1-800-456-1477.
Initiated
2019-03-01
Root cause
Device Design
Terminated
2022-08-10
Distribution
US Nationwide distribution in the states of MN, WI, TX, NC, GA, OK, UT, NY, SC, IL, LA, MI, IN, CA, OH, WA. Foreign (OUS): AUSTRIA, BELGIUM, CANADA, ESTONIA, FRANCE, GERMANY, GREECE, INDIA, ITALY, KUWAIT, LIECHTENSTEIN, LITHUANIA, NETHERLANDS, POLAND, SAUDI ARABIA, SINGAPORE, SPAIN, SWITZERLAND, TURKEY, UNITED KINGDOM
Recall number
Z-1203-2019
Recall status
Terminated
CDRH record ID
171649
FDA classified
2019-04-19
Recall event number
82458

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗