Open original FDA record ↗Recall record
- FDA identifier
- Z-1203-2019
- Company in FDA record
- Thoratec Corp.
- FDA classification date
- Apr 19, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Small black plastic particle was observed in the left ventricle during implant procedure.
Source facts
- Action
- On 03/01/19, sales representatives started visiting and telephoning consignees informing them to quarantine affected devices. The firm requested that consignees complete and return the Device Retrieval Template. On 03/08/19, Urgent Medical Device Recall notices were hand delivered to all impacted accounts. Customers were advised to stop use, quarantine, and return any units on hand. For patients, who's implant procedure included the use of affected devices, firm recommends close vigilance for potential harm, which could include concerns such as infection or thromboembolism. The coring knife is not impacted. Firm requests that customers complete and return the acknowledgement form. Firm recommends that customer with additional questions call 1-800-456-1477.
- Initiated
- 2019-03-01
- Root cause
- Device Design
- Terminated
- 2022-08-10
- Distribution
- US Nationwide distribution in the states of MN, WI, TX, NC, GA, OK, UT, NY, SC, IL, LA, MI, IN, CA, OH, WA. Foreign (OUS): AUSTRIA, BELGIUM, CANADA, ESTONIA, FRANCE, GERMANY, GREECE, INDIA, ITALY, KUWAIT, LIECHTENSTEIN, LITHUANIA, NETHERLANDS, POLAND, SAUDI ARABIA, SINGAPORE, SPAIN, SWITZERLAND, TURKEY, UNITED KINGDOM
- Recall number
- Z-1203-2019
- Recall status
- Terminated
- CDRH record ID
- 171649
- FDA classified
- 2019-04-19
- Recall event number
- 82458