Z-1203-2021: Omnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX

Recall record Z-1203-2021 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1203-2021
Company in FDA record
Insulet Corporation
FDA record creation date
Mar 10, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended

Source facts

Action
Insulet issued Email, mail, or outbound call notification of the Field Safety Notification between Feb 2nd and Feb 5 th to end customers, HCPs and distributors in: Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland. Letter states reason for correction, health risk and action to take: Please refer to the attached Annex in reference to page 6 of the User Guide as opposed to referring to the instructions in the existing User Guide. We recommend you place this insert at p.6 in your User Guide. Please ensure that the comma is always used when entering bolus values. In accordance with applicable rules, this corrective action is implemented with the knowledge of the relevant competent authorities. Please electronically confirm your understanding of this Field Safety Notification by going to the webpage the link below and following the instructions on screen.
Initiated
2021-02-01
Root cause
Labeling design
Terminated
2023-02-08
Distribution
Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland
Recall number
Z-1203-2021
Recall status
Terminated
CDRH record ID
185770
FDA record created
2021-03-10
Recall event number
87305

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗