Open original FDA record ↗Recall record
- FDA identifier
- Z-1203-2026
- Company in FDA record
- Cardinal Health 200, LLC
- FDA classification date
- Jan 28, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm is updating the IFU to clarify the intended patient population for the identified chest drainage units and accessories as adults, 18 years and older. When used on infants, the CDU may not demonstrate tidaling/bubbling, which may confound the clinician's assessment of device functionality and lead to inadequate/inappropriate treatment/therapy, delay to treatment/therapy, and prolonged hospitalization.
Source facts
- Action
- On December 26, 2025, the firm notified customers via Urgent Medical Device Product Advisory letters. Customers were instructed to continuing utilizing the CDUs on adult patients (not pediatric patients or infants). The updated IFU for affected product codes can be found at https://www.mycardinalmsds.com/
- Initiated
- 2025-12-26
- Root cause
- Error in labeling
- Distribution
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, Chile, Colombia, Dominican Republic, and Mexico.
- Recall number
- Z-1203-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217947
- FDA classified
- 2026-01-28
- Recall event number
- 98228