Z-1204-2023: Med-Riser MR600 Patient Lift

Recall record Z-1204-2023 for Med-Mizer, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1204-2023
Company in FDA record
Med-Mizer, Inc.
FDA classification date
Mar 2, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Risk of boom pivot failing due to the nut securing the boom becoming loose and the bolt to be unsecure to the flange on the side.

Source facts

Action
Med-Mizer notified distributors of this recall event by distributing an Urgent Medical Device Recall Notice on 02/03/2022 via Certified Mail. The MR600 and STS500 Patient Lifts are being subject to recall due to the potential for the nut securing the boom to become loose over time, which could result in the boom moving unexpectedly side to side or dropping the patient during operation. Customers are asked to inspect all affected devices at their facility to determine if the nut securing the boom is loose. If it is, immediately remove the device from service. Med-Mizer is providing customers with affected devices replacement parts in order to complete the correction. Customers are to complete the provided Correction Response Form to Med-Mizer and return via fax at (812)932-5678, by email at support@med-mizer.com, or by mail at 80 Commerce Drive, Batesville, IN 47006. If your firm is a distributor, please provide this notice to your customers. Any questions can be directed to Med-Mizer by phone at (877)867-7365.
Initiated
2023-02-03
Root cause
Device Design
Distribution
Awaiting consignee list from the recalling firm.
Recall number
Z-1204-2023
Recall status
Open, Classified
CDRH record ID
198550
FDA classified
2023-03-02
Recall event number
91695

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗