Firm became aware that a potential pinch point can exist between the stop level and the safety catch when rotating the x-ray tube assembly. Risk of injury to operators fingers.
Source facts
Action
The recalling firm has issued a field correction per Update Instructions AX012/06/S. This update instructs a Siemens Customer Service Engineer to visit affected consignees to install a protective cover and additional warning labels on the device, beginning 04/28/2006.
Initiated
2006-04-28
Root cause
Other
Terminated
2008-09-30
Distribution
The products were shipped nationwide to medical facilities.
Recall number
Z-1205-06
Recall status
Terminated
CDRH record ID
45995
FDA classified
2006-07-06
Recall event number
35395