Open original FDA record ↗Recall record
- FDA identifier
- Z-1205-2019
- Company in FDA record
- Brainlab AG
- FDA record creation date
- Apr 21, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is a potential for an incorrect dose distribution calculation by Brainlab RT Elements software (for affected versions) under specific circumstances when using the Pencil Beam algorithm on the GPU (graphics card), as is the default system setting.
Source facts
- Action
- On March 11, 2019, Brainlab sent FIELD SAFETY NOTICE/PRODUCT NOTIFICATION emails to their consignees advising them of the problem. Brainlab will disable the Pencil Beam GPU calculation in the settings of the affected RT Elements Radiation Treatment Planning Software, and thus correct the described issue permanently for all affected customers.
- Initiated
- 2019-03-11
- Root cause
- Device Design
- Terminated
- 2021-05-12
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IN, KY, LA, MA, MD, MS, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, & WV. Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, El Salvador, Finland, France, Germany, Hong Kong, Iran, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Martinique, Mexico, Monaco, Morocco, Netherlands, Poland, Portugal, Russian Fed., South Africa, South Korea, Spain, Switzerland, Taiwan, Tunisia, United Arab Emirates, and United Kingdom.
- Recall number
- Z-1205-2019
- Recall status
- Terminated
- CDRH record ID
- 172010
- FDA record created
- 2019-04-21
- Recall event number
- 82601