Skip to content

Public FDA record

Z-1205-2022: Proteus235- Proton Therapy System: to produce and deliver a proton beam for the treatment of patients wi

Recall record Z-1205-2022 for Ion Beam Applications S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1205-2022
Company in FDA record
Ion Beam Applications S.A.
FDA classification date
May 27, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

First layer of Uniform Scanning treatment fields is sometimes irradiated with incorrect Scanning Magnet setpoints resulting in the field size may be smaller than expected and the irradiated area may receive more dose than prescribed

Source facts

Action
***UPDATE 12/22/22*** Ion Beam Application S.A. (IBA S.A.) issued an Amended Urgent Medical Device Correction to customers starting December 22, 2022 (supersedes previous letter Sept 15, 2022). Letter provides quantitative data about the impact on the dose distribution, and the description of the solution (software upgrade) as well as the time schedule for its deployment. Cyclic check thresholds verification and software upgrade Additionally, IBA will provide each impacted customer with a report including: - The list of the Uniform Scanning fields with the first layer entirely delivered with the tuning setpoints and a significant difference (above 0.3V) between the prescribed setpoints and the tuning setpoints, - The modelled maximum and average dose errors for these fields IBA will upgrade the PTS version to ensure that: - the irradiation of a layer cannot be started if the Scanning Magnet setpoints for the layer are not properly sent to the Scanning Magnet Power Supply, and - no setpoints request can be lost (even if previous requests are not yet executed or completed). The report including the identified impacted Uniform Scanning fields with the modelled maximum and average dose errors for these fields will be provided to each impacted customer by February 28, 2023. On the sites where IBA has not yet finalized the verification and the possible adaptation of the field size and dose rate cyclic check thresholds, this action will be completed by January 15, 2023. The PTS version upgrade will be deployed on your site by March 31, 2023. Contact: HEAD OF POST MARKET VIGILANCE Sonia PINEL Vigilance@iba-group.com +32 10 497 516 ___________________________________________________________________________ **UPDATED 9/15/22*** The cause of these events has been identified as a software race condition between the different tasks that handle the Scanning Magnet Electronic Unit (SMEU). IBA CAPA PNC-543 has been opened. The root cause analysis is in progress. This recall re
Initiated
2022-04-20
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of IL, VA, WA and the country of Germany.
Recall number
Z-1205-2022
Recall status
Open, Classified
CDRH record ID
193836
FDA classified
2022-05-27
Recall event number
90213

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.