Open original FDA record ↗Recall record
- FDA identifier
- Z-1206-2023
- Company in FDA record
- Carestream Health, Inc.
- FDA classification date
- Mar 2, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Source facts
- Action
- Carestream issued Urgent Medical Device Correction Letter to Director of Radiology, Chief Radiology Administrator and DRX-Compass/DR-FIT X-ray System Administrators (US) consignees on 1/30/23. Letter states reason for recall, health risk and action to take: Do not use the Z-Axis Motorized button on the Tube Head Display at this time. Use the Brake Release buttons (Items 5 or 8) to manually move the OTC in the Z-Axis. " Press any E-stop button to stop unexpected movement immediately. " A Service Representative will install a Firmware correction. Action to be taken: A Carestream Service Engineer or Carestream Health Authorized Representative will contact your site and setup a service call to inspect your device and install a new firmware version. For your convenience, a Notification Acknowledgement form has been provided with this letter. Please read and complete the information within 5 working days. Remit the completed form by e-mail to: postmarketra@carestream.com. Questions or concerns, please contact the Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24 hour basis. Outside of the U.S., please call your local Carestream S vice support number
- Initiated
- 2023-01-30
- Root cause
- Software design
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
- Recall number
- Z-1206-2023
- Recall status
- Open, Classified
- CDRH record ID
- 198578
- FDA classified
- 2023-03-02
- Recall event number
- 91698