Skip to content

Public FDA record

Z-1207-2018: 6F Taiga Guiding Catheter

Recall record Z-1207-2018 for Medtronic Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1207-2018
Company in FDA record
Medtronic Inc
FDA classification date
Mar 29, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Complaints regarding 6F Taiga guide catheter tip detachment and/or tip splitting.

Source facts

Action
A recall notification was sent to all costumers in September 2017. Customers actions to be done: Identify and quarantine all unused affected products as listed in your inventory. Share notification with others who need to be aware where affected product may have potentially been transferred.
Initiated
2017-09-26
Root cause
Nonconforming Material/Component
Terminated
2021-04-27
Distribution
Japan
Recall number
Z-1207-2018
Recall status
Terminated
CDRH record ID
162346
FDA classified
2018-03-29
Recall event number
79449

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗