Z-1207-2022: HeartWare HVAD Pump Implant Kit, a) Model Number 1104JP, b) Model Number MCS1705PU

Recall record Z-1207-2022 for Medtronic Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1207-2022
Company in FDA record
Medtronic Inc
FDA record creation date
Jun 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The pump may have a welding defect that can lead the pump to malfunction. The issue is under investigation by the firm. It is suspected that a weld defect on the center post cap may allow moisture into the center post and corrode the magnets that keep the impeller rotating concentrically, leading to unintended contact between the center post and impeller. If this occurs, it may affect blood flow through the pump. The issue may present clinical signs and symptoms that resemble pump thrombosis.

Source facts

Action
The firm issued letters flagged "Urgent Medical Device Recall" dated April 2022 on 4/14/2022 via 2-day UPS delivery. The letter provided the issue description and requested the consignee to take the following actions: Immediately quarantine and return the unused affected product, complete the enclosed Customer Confirmation Form and email it to the firm, share the notice with all those who need to be aware within their organization or to any organization where the existing product has been transferred, and retain the notification for their records regarding the retrieval of the unused affected product. The Customer Confirmation Form confirms the consignee has read the letter and taken appropriate action. The consignee is to indicate the product and serial number that is being returned. The consignee is also supposed to list the product and serial number of the HVAD Pump implant kits that have been implanted. Return instructions are also provided. Beginning April 26, 2022, the firm send a follow-up communication to all VAD centers with patients on support to communicate that this retrieval has been conducted as well as informing them of the ongoing investigation and providing patient management recommendations.
Initiated
2022-04-14
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide DE and TX. There was no military/government distribution. The countries of Germany and Japan.
Recall number
Z-1207-2022
Recall status
Open, Classified
CDRH record ID
192998
FDA record created
2022-06-03
Recall event number
89966

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗