Z-1207-2025: BD Pyxis MedFlex 2.0 REF 1139-00 Bio-ID Spare Parts REF 353200-01 The BD Pyxis" MedFlex is an Automated

Recall record Z-1207-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1207-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Feb 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to fingerprint scanner failing resulting in the scanner heating up.

Source facts

Action
On 1/23/2025, the firm sent a "Urgent Medical Device Correction" Letter via FedEx and email to customers informing them that the firm became aware via customer complaints that an internal component within the Bio-ID (fingerprint scanner) is failing and resulting in the Bio-ID scanner heating up and could potentially cause superficial burns if finger is left on the scanner for up to or more than 5 seconds. Customers are instructed: 1. BD Pyxis" SupplyStation" and MedFlex 2.0 can continue to be used as intended. If you suspect that the Bio-ID Scanner is hot, please discontinue the use of the Bio-ID Scanner, submit a complaint to BD, and use your username and password to login instead. 2. Please return the customer response form to acknowledge receipt of this notice. For questions/assistance: -Technical Support - Phone: 1-866-930-9251 Phone hours: 24 hours, 7 days a week for Technical Questions. -Regional Complaint Center - Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted. Mon-Fri 8:00am-5:00pm CT or Email: productcomplaints@bd.com for Product Complaints -Field action related questions - Email (pending) - Email for receipt of customer response forms and recall related questions.
Initiated
2025-01-23
Root cause
Nonconforming Material/Component
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, Guam, HI , IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, and WY. The countries of Australia, Bahrain, Belgium, Brazil, Canada, Hong Kong, Kuwait, Lebanon, New Zealand, Qatar, Saudi Arabia, Spain, and Thailand.
Recall number
Z-1207-2025
Recall status
Open, Classified
CDRH record ID
211734
FDA classified
2025-02-21
Recall event number
96004

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗