Skip to content

Public FDA record

Z-1208-03: Software Versions V1.2R,or earlier for the Harvard 2, Dual Infusion Pump (P/N 2002-001)

Recall record Z-1208-03 for Harvard Clinical Technology. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1208-03
Company in FDA record
Harvard Clinical Technology
FDA classification date
Sep 11, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Under specific sequence conditions, the unit may apper to be infusing, but is not

Source facts

Action
Harvard Medical Technology inc. notified customers by letter on 8/11/03, advising users of the software upgrade to version V1.3T. Facilities will be visited by technicians who will upgrade the software.
Initiated
2003-08-11
Root cause
Other
Terminated
2012-05-07
Distribution
Nationwide
Recall number
Z-1208-03
Recall status
Terminated
CDRH record ID
28708
FDA classified
2003-09-11
Recall event number
26903

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗