Z-1208-2022: Resin UPDATE CD-ROM, REF: 250-3020 inside VARIANTnbs Sickle Cell Program Reorder Pack, REF: 250-3000

Recall record Z-1208-2022 for Bio-Rad Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1208-2022
Company in FDA record
Bio-Rad Laboratories, Inc.
FDA record creation date
May 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The problem is with the Bio-Rad VARIANT nbs Sickle Cell Program Resin Update CD-ROM software; Model Number: 250-3020, included in VARIANT nbs Sickle Cell Progam Reorder Pack #250-3000. This CD-ROM software included in specific lots of VARIANT nbs Sickle Cell Progam Reorder Pack causes all customized settings in the Setup/Test/Data Setup subscreen and Setup/Test/Pattern Setting subscreen to be overwritten with the default settings when the Update Kit procedure is performed.The VARIANTnbs Sickle Cell Program has been validated using the default Resin Update CDROM parameters for the intended use. However, there is a risk that the customer's custom pattern settings may be different than the default settings, resulting in a different pattern being assigned to some samples that the laboratory has validated. Following the complaints from 4 customers who experienced these issues, the firm initiated this recall.

Source facts

Action
On 04/07/22, Field Action Notifications were sent to customers, who were informed of the following. Not affected by this action if A) affected Resin Update CD-ROM was not installed. B) Default settings are used (i.e., did not edit any settings) in the Setup/Test/Data Setup and/or Pattern Setting subscreens prior to performing the Update Kit procedure using one of the affected CD-ROMs. Affected by this action if customized settings were used and subsequently the Update Kit procedure using one of the affected CD-ROMs was performed; your customized settings were overwritten by the default Bio-Rad settings. If affected: 1. Go to the Setup/Test/Data Setup and/or Setup/Test/Pattern Setting subscreen. Click the Allow Editing button and enter the password. 2. Enter your preferred settings in each of the applicable subscreens. 3. Click the Lock Editing button. 4. Do not perform the Update Kit procedure again with any of the affected Resin Update CD-ROMs. 5. Review all sample reports specifically for your customized setting(s) that reverted to the default setting(s). Refer to Operation Manual (LB001682revB) Section 6.1.7 Additionally, If your lab customized selections in the Data Setup tabs (Sample Types, Peak Result Export, and/or Sample Result Export), then reset them. The default settings have all pertinent patient information selected; any unselected parameters do not affect the patient results. LIS not affected; refer to Software Operation Manual Section 6.1.6. If assistance is needed with recovering overwritten customized settings, an old database can be temporarily restored and then the customized Data Setup settings and/or Pattern settings can be copied/printed. Restore the working database before proceeding to prevent loss of data. Complete and return the response form. The following contact was provided by the firm: 510-724-7000
Initiated
2022-04-07
Root cause
Packaging process control
Distribution
US: AZ, AR, GA, CO, TN, LA, VA, FL, KY, IN, KS, MD, MN, MO, NC, NJ, NY, OH, OR, SC, IA, AL, UT, PR OUS: Canada, Spain, France, Italy, India, Vietnam, Philippines
Recall number
Z-1208-2022
Recall status
Open, Classified
CDRH record ID
193175
FDA record created
2022-05-28
Recall event number
90027

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗