Open original FDA record ↗Recall record
- FDA identifier
- Z-1208-2023
- Company in FDA record
- Coltene Whaledent Inc
- FDA classification date
- Mar 2, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Screw connection between the turbine head and the push button may loosen and fall off during the treatment, and increase the risk of swallowing or inhaling part of the turbines head or the push button
Source facts
- Action
- Coltene issued Urgent Medical Device Correction to Distributors. The written notification included the Field Safety Notice issued by the manufacturer and the Correction Notice that was issued on January 20th, 202 via Email. Letter states reason for notification, health risk, and action to take: Warning: When using the STATIS SL turbines with REF 2.000.7300 / S10001, with the lack of care or age-related wear and tear, there is a possibility that the push button on the turbine will come loose and this will increase the risk of swallowing or inhaling part of the turbines head or the push button. Recommended action for users: Please urgently observe the following safety instructions from the instructions for use on handling, care and running behaviour of the turbine to protect your patients. This notice needs to be passed on to all those who need to be aware, both within your own organization and in organizations where the potentially affected devices have been transferred. Please ensure that this information and the associated action are followed for a reasonable period of time to ensure the effectiveness of the action.
- Initiated
- 2023-01-20
- Root cause
- Other
- Distribution
- Nationwide Foreign: Canada Germany Indonesia Ireland Lebanon Saudia Arabia Turkey
- Recall number
- Z-1208-2023
- Recall status
- Open, Classified
- CDRH record ID
- 198315
- FDA classified
- 2023-03-02
- Recall event number
- 91601