A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Source facts
Action
On 2/5/2025, correction notices were emailed to customers who were asked to do the following:
The following work around can be completed for any impacted SUV calculation. The correct SUV values
may be obtained in the SUV calculator, as follows:
1. Adjust the administration time and series time in IntelePACS consistent with the actual time of
injection and the time of series acquisition.
2. Verify that the calculated elapsed time is correct.
3. Recalculate the SUV in the calculator, this is the correct SUV value.
- Customers are asked to review scans and notify the recalling firm if scans were performed on Daylight Savings Time dates since 2022.
- A software upgrade to eliminate this bug is expected to be released by the end of March 2025. Once available, you will be notified via email. You will then be able to schedule your product upgrade.
- Complete and return the acknowledgement form via email to regulatory-affairs@intelerad.com
For more information or to ask any questions about this notification, please contact the firm's regulatory affairs department at regulatory-affairs@intelerad.com
Initiated
2025-02-05
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Recall number
Z-1208-2025
Recall status
Open, Classified
CDRH record ID
212457
FDA classified
2025-02-21
Recall event number
96270
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.