Z-1209-2019: Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1.

Recall record Z-1209-2019 for Fenwal Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1209-2019
Company in FDA record
Fenwal Inc
FDA record creation date
Apr 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the centrifuge tubing line making contact with the washing chamber. The contact may result in excessive wear of the centrifuge tubing or in extreme cases a puncture of the tubing.

Source facts

Action
The initial distributor was notified by Fresenius Kabi on 03/05/2019, who provided their distribution list. Fresenius Kabi notified those customers on about 03/11/2019 via "IMPORTANT - CATSmart Software Upgrade and AT-1 sets used on CATSmart" letter. Customers were instructed that Fresenius Kabi will be implementing a software changed to improve priming acceleration. As an interim solution, Fresenius Kabi instructed that CATSmart users should switch to using AT-3 Autotransfusion sets and kits: Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits (AT-3). Additionally customers were requested to complete and return the Customer Reply Form. Customers will be contacted by a Fresenius Kabi service representative to schedule a service appointment for updating the software. They anticipate changing the software by October 1, 2019. Questions can be directed to Fresneius Kabi at 800-333-6925.
Initiated
2019-03-05
Root cause
Device Design
Terminated
2020-10-05
Distribution
Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and PR, and country of Canada.
Recall number
Z-1209-2019
Recall status
Terminated
CDRH record ID
171537
FDA record created
2019-04-23
Recall event number
82395

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗