Open original FDA record ↗Recall record
- FDA identifier
- Z-1209-2019
- Company in FDA record
- Fenwal Inc
- FDA record creation date
- Apr 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the centrifuge tubing line making contact with the washing chamber. The contact may result in excessive wear of the centrifuge tubing or in extreme cases a puncture of the tubing.
Source facts
- Action
- The initial distributor was notified by Fresenius Kabi on 03/05/2019, who provided their distribution list. Fresenius Kabi notified those customers on about 03/11/2019 via "IMPORTANT - CATSmart Software Upgrade and AT-1 sets used on CATSmart" letter. Customers were instructed that Fresenius Kabi will be implementing a software changed to improve priming acceleration. As an interim solution, Fresenius Kabi instructed that CATSmart users should switch to using AT-3 Autotransfusion sets and kits: Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits (AT-3). Additionally customers were requested to complete and return the Customer Reply Form. Customers will be contacted by a Fresenius Kabi service representative to schedule a service appointment for updating the software. They anticipate changing the software by October 1, 2019. Questions can be directed to Fresneius Kabi at 800-333-6925.
- Initiated
- 2019-03-05
- Root cause
- Device Design
- Terminated
- 2020-10-05
- Distribution
- Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and PR, and country of Canada.
- Recall number
- Z-1209-2019
- Recall status
- Terminated
- CDRH record ID
- 171537
- FDA record created
- 2019-04-23
- Recall event number
- 82395