Skip to content

Public FDA record

Z-1211-03: Exactech Tibial Augmentation Block Trial Size 0, Exactech Tibial Tray Trial size 1 Delta and Exactech Tra

Recall record Z-1211-03 for Exactech, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1211-03
Company in FDA record
Exactech, Inc.
FDA classification date
Sep 11, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Misalignment of the fixation holes between the tibial augment block trials and the corresponding tibial tray trials and trapezoid tray trials. The misalignment of the pin fixation holes only creates an issue when augmentation is used.

Source facts

Action
A recall letter dated July 31, 2003, requesting return of certain Tibial Augmentation Block Trials and an advisory (Field Advisory K-078) on the use of certain Tibial Tray Trials and Trapezoid Tibial Tray Trials. The recall letter gives the address to return product as: Exactech, 2243 NW 66th Court, Gainesville, FL. Consignees were told to immediately cease distribution or use of the recalled products and to return them. Also the letter asks them to extend the recall information to their user accounts and to read and comply with the instructions provided in the attached 'Field Advisory K-078'. This advisory instructs the dealers and users on the use of the devices to avoid problems.
Initiated
2003-08-01
Root cause
Other
Terminated
2012-05-03
Distribution
Nationwide. Also the products were distributed to three intenational accounts in Spain, Japan and China.
Recall number
Z-1211-03
Recall status
Terminated
CDRH record ID
28747
FDA classified
2003-09-11
Recall event number
26919

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗