Open original FDA record ↗Recall record
- FDA identifier
- Z-1211-03
- Company in FDA record
- Exactech, Inc.
- FDA classification date
- Sep 11, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Misalignment of the fixation holes between the tibial augment block trials and the corresponding tibial tray trials and trapezoid tray trials. The misalignment of the pin fixation holes only creates an issue when augmentation is used.
Source facts
- Action
- A recall letter dated July 31, 2003, requesting return of certain Tibial Augmentation Block Trials and an advisory (Field Advisory K-078) on the use of certain Tibial Tray Trials and Trapezoid Tibial Tray Trials. The recall letter gives the address to return product as: Exactech, 2243 NW 66th Court, Gainesville, FL. Consignees were told to immediately cease distribution or use of the recalled products and to return them. Also the letter asks them to extend the recall information to their user accounts and to read and comply with the instructions provided in the attached 'Field Advisory K-078'. This advisory instructs the dealers and users on the use of the devices to avoid problems.
- Initiated
- 2003-08-01
- Root cause
- Other
- Terminated
- 2012-05-03
- Distribution
- Nationwide. Also the products were distributed to three intenational accounts in Spain, Japan and China.
- Recall number
- Z-1211-03
- Recall status
- Terminated
- CDRH record ID
- 28747
- FDA classified
- 2003-09-11
- Recall event number
- 26919