Open original FDA record ↗Recall record
- FDA identifier
- Z-1211-2019
- Company in FDA record
- COVIDIEN LLC
- FDA record creation date
- Apr 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sterilization method used was not consistent with the labeling and the approved sterilization method
Source facts
- Action
- Medtronic notified consignees via an "URGENT MEDICAL DEVICE RECALL" letter by Federal Express and/or certified mail on March 13, 2019. The letter informs customers of the recall of twenty-one production lot of Covidien Endo Clinch" and Endo Grasp" Auto Suture" graspers 5mm as the sterilization method used for these lots was not consistent with the labeling and the approved sterilization method. Customers are instructed to take the following actions: 1. Please immediately quarantine and discontinue use of the affected item codes and lots listed above. 2. Please return affected product as indicated below. 3. If you have distributed the Endo Clinch" II or Endo Grasp" Auto Suture" graspers 5mm listed above, please promptly forward the information from this letter to those recipients. 4. Complete the Recalled Product Return Form even if you do not have inventory. -----If you have inventory email to: feedback.customerservice@covidien.com fax to: 800-895-6140. -----Exception: Customers with zero inventory, fax to 203-492-7719 or email to GMBFCAMITG@Medtronic.com If purchased by distributor, fax to 203-492-7719 or email GMBFCAMITG@Medtronic.com. If you have any questions or concerns, please do not hesitate to contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
- Initiated
- 2019-03-13
- Root cause
- Process control
- Terminated
- 2020-04-23
- Distribution
- Worldwide Distribution: US (Nationwide) and to countries of: Canada, Chile, Costa Rica, Guatemala, Japan, Mexico, Panama, Slovakia, Spain, Switzerland and Turkey.
- Recall number
- Z-1211-2019
- Recall status
- Terminated
- CDRH record ID
- 171603
- FDA record created
- 2019-04-23
- Recall event number
- 82426