Z-1211-2022: Drager SafeStar 55, Catalog No. MP01790

Recall record Z-1211-2022 for Draeger Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1211-2022
Company in FDA record
Draeger Medical, Inc.
FDA record creation date
Jun 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Possible occlusion of filters due to manufacturing error.

Source facts

Action
The firm distributed a notification titled "URGENT MEDICAL DEVICE RECALL" by mail on 05/16/2022, in which it informed customers of the voluntary recall of Lot LT2103 of Draeger SafeStar 55 Breathing System Filters, Ref. No. MP01790, UDI-DI 04048675026785. The firm states that this recall is being initiated in response to becoming aware of a complaint in which a patient became hypoxic and had to be resuscitated while under anesthesia using the Draeger SafeStar 55 Breathing System Filter from the affected lot. The firm indicated that a manufacturing feature that automatically destroys filters that fail acceptance testing was disabled during servicing, which resulted in filters having to be hand sorted during manufacturing of Lot LT2103. Drager is asking customers to check their inventory for SafeStar 55 breathing system filters from Lot LT2103. If any affected product is on hand, customers are to call Draeger Customer Service at 1-80-437-2437, fax at 1-215-723-5935, or email at US-Medical@draeger to obtain free replacement of product. If any affected product is on hand, customers are to segregate and quarantine product until Draeger's provision of a return materials authorization and pre-paid shipping labels. Draeger also asks customers to make the recall notification available within their organization and to forward a copy if product was further distributed. A Customer Acknowledgement Form was provided with the notification for all customers to complete and return. Customers with any questions are to contact Michael Kelhart, Quality Director, between the hours of 8:00 AM and 4:30 PM EST at 1-800-437-2437, press 1, then 32349.
Initiated
2022-05-16
Root cause
Process design
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, NC, ND, NV, NY, OH, PA, TN, TX, WA, and WV. The countries of Austria, Belgium, Bulgaria, Canada, Columbia, Croatia, Czech Republic, Denmark, France, Germany, India, Indonesia, Ireland, Italy, Ivory Coast, Japan, Kazakhstan, Latvia, Malaysia, Morocco, Netherlands, Norway, Philippines, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Spain, Suriname, Sweden, Switzerland, Thailand, United Kingdom, and United Arab Emirates.
Recall number
Z-1211-2022
Recall status
Open, Classified
CDRH record ID
193712
FDA record created
2022-06-10
Recall event number
90192

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗