Z-1211-2023: Sprint 200 wheeled stretcher, Item numbers 1ES200-7, 1ES210-1, and 1ES211-1.

Recall record Z-1211-2023 for Linet Spol. S.r.o.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1211-2023
Company in FDA record
Linet Spol. S.r.o.
FDA classification date
Mar 4, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling being updated to assure user is securing the siderail to a locked position to prevent potential patient fall and injury.

Source facts

Action
The firm issued "URGENT MEDICAL DEVICE CORRECTION" recall letters dated 1/17/2023 via email on 1/18/2023. The letter explains the issue that an audible clicking sound can be heard when the side rail is raised, but this sound does not indicate the side rail has locked into the raised position. The consignee is to take immediate action to notify caregivers using the Sprint 200 of this potential hazard and proper use for securing a side rail in the raised position. The consignee is informed that Linet will be adding additional safety warning labels located at each side rail latch (2 per side of stretcher), add additional safety warning labels to both side rails (1 per side of stretcher), and that additional training material and an updated copy of the Sprint 200 Emergency Stretcher Instructions for Use and Technical Description manual is available using the QR code in the letter or by following a link to access the materials. The firm's service department will be contacting the consignees to schedule an appointment to perform the corrections. Photos of the additional warning labels are included in the letter. Please contact LINET service department at 877-815-9897 with immediate questions or concerns.
Initiated
2023-01-18
Root cause
Labeling design
Distribution
US Distribution was made to AL, CA, FL, IA, KY, LA, MN, NV, NC, OK, TX, VA, WA, and WV. There was no government/military distribution. Foreign distribution was made to Australia, Bahrain, Belgium, Germany, Great Britain, Italy, Netherland, Singapore, South Africa, Switzerland, and United Emirates Canada.
Recall number
Z-1211-2023
Recall status
Open, Classified
CDRH record ID
198058
FDA classified
2023-03-04
Recall event number
91492

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗