Skip to content

Public FDA record

Z-1212-2022: XVIVO Organ Chamber REF 19020

Recall record Z-1212-2022 for XVIVO PERFUSION AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1212-2022
Company in FDA record
XVIVO PERFUSION AB
FDA record creation date
Jun 1, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The weld seal of the primary pouch was incomplete, and the sterility of the product cannot be ensured/guaranteed.

Source facts

Action
On or about 04/11/2022, XVIVO Inc. issued an "URGENT- Field Safety Notice" for its XVIVO Organ Chamber via email to customers informing them XVIVO has identified that the primary package for a number of batches may have notable defects in the weld of the bag. The notable defects are unsealed sections of the weld (i.e. the bag is not completely closed). Unsealed sections in the weld may compromise the sterility of the product and therefore it is important to NOT use a product showing this defect. Customer are instructed to: 1. Carefully check the primary package for unsealed sections. A follow-up "URGENT:-Medical Device Correction" issued on or about 05/06/2022, this letter provides photos of example of unsealed welds. 2. If unsealed sections are noted, do not use the product and contact XVIVO Customer Support at customersupport@xvivogroup.com for further action. 3. if not defects are noted, the product may be used as intended. Any questions or additional information, contact Customer Support at customersupport@xvivogroup.com or telephone number +1-303-395-9171.
Initiated
2022-04-08
Root cause
Under Investigation by firm
Terminated
2025-01-30
Distribution
U.S.: CA, FL, MA, NC, NY, OH, PA, and TX O.U.S.: Austria, Canada, France, Ireland, Italy, and the Netherland
Recall number
Z-1212-2022
Recall status
Terminated
CDRH record ID
193567
FDA record created
2022-06-01
Recall event number
90143

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗