Z-1213-2023: Baxter MiniCap Disconnect Cap with Povidone-Iodine Solution, 5C4466P; Accessory to disposable administrat

Recall record Z-1213-2023 for Baxter Healthcare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1213-2023
Company in FDA record
Baxter Healthcare Corporation
FDA classification date
Mar 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

MiniCap Disconnect Cap with Povidone-Iodine are packaged in foil pouches, which may have been incorrectly sealed, i.e., the pouches may have open or weak seals. This could lead to exposure to air, resulting in insufficient iodine/dry sponge inside the MiniCap, which could lead to the potential for inadequate disinfectant.

Source facts

Action
Baxter Healthcare initiated the recall by issuing an Urgent Medical Device Recall notice to its consignees by letter on 02/01/2023. The notice requested the following from customers: 1. Check your stock and set aside all product with the affected lot numbers. 2. Contact Baxter HomeCare Services to arrange for return of impacted product and for ordering replacement product. 3. If all of your product is impacted by this recall and you are unable to get unaffected lot numbers, resulting in a significant delay in therapy, please follow the below steps: a. Contact your clinic for clinical guidance and check each pouch and do not use the MiniCap contained inside of any pouches that appear to have damaged seals. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday." Recall was expanded and Baxter Healthcare sent an updated notification to customers and patients on 05/09/2023. The letter instructed consignees and patients to immediately cease use of the affected product, arrange for the return of affected product and assess each patient's individual next steps based on clinical evaluation. Customers were also instructed to complete the customer response form via the portal (for products directly received from Baxter) or through the distributor. If the products were further distributed, please notify customers.
Initiated
2023-02-01
Root cause
Process control
Distribution
US Nationwide distribution and Canada.
Recall number
Z-1213-2023
Recall status
Open, Classified
CDRH record ID
198353
FDA classified
2023-03-03
Recall event number
91612

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗