Open original FDA record ↗Recall record
- FDA identifier
- Z-1214-2010
- Company in FDA record
- Medtronic Neuromodulation
- FDA classification date
- Mar 29, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Routine monitoring of Medtronic's vigilance program identified user related programming errors. Medtronic has received reports of both overdose and underdose that were the result of programming use of SynchroMed implantable infusion pumps. These errors were most commonly associated with: --Priming Bolus Programming; and --Bridge Bolus Programming. Pump programming errors that result in overdos
Source facts
- Action
- Medtronic is exchanging 8870 versions AAG, AAJ and BBG and replacing it with the updated versions AAH01 and BBH01. The updated versions contain user interface improvements to reduce the chance of user related programming errors when programming SynchroMed II and SynchroMed EL Infusion Pump system. A Medtronic "Medical Device Correction" letter was left behind for consignees. The letter was addressed to "Healthcare Provider" and was dated February 2009. The letter provided information regarding changes to the programming software used in SynchorMed Infusion System. For additional information, please contact your Medtronic Representative or Medtronic Neuromodulation Technical Services at 1-800-707-0933.
- Initiated
- 2009-03-23
- Root cause
- Other
- Terminated
- 2012-07-10
- Distribution
- All States in the US including DC. OUS: Austria Belgium Croatia Cyprus Czech Republic Denmark Estonia Faroe Islands Finland France French Polynesia Germany Greece Hungary India Ireland Israel Italy Kuwait Lebanon Malta Martinique Netherlands New Caledonia Norway Poland Portugal Reunion Russian Federation San Marino Saudi Arabia South Africa Spain Sweden Switzerland Turkey United Kingdom Vatican City State Canada Mexico Brasil
- Recall number
- Z-1214-2010
- Recall status
- Terminated
- CDRH record ID
- 88556
- FDA classified
- 2010-03-29
- Recall event number
- 54477