- FDA identifier
- Z-1214-2012
- Company in FDA record
- Spectrum Dynamics
- FDA classification date
- Mar 21, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The firm received a report related to the detector housing on the D-SPECT Cardiac Scanner. Specifically, it was reported that if the detector housing securing mechanism is compromised, the detector housing may pull away from its secured position on the gantry and potentially come into contact with a patient. The reported event was discovered during set up and thus did not involve a patient or p
Source facts
- Action
- Spectrum Dynamics sent a "Voluntary Field Correction" letter dated July 15, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken. To implement this fix, a BC Technical Field Service Engineer contacted users to schedule a service visit. In the meantime, Spectrum Dynamics did not believe that this issue warranted the discontinuation of use of the D-SPECT Cardiac Scanner. However, in an abundance of caution, if the detector housing appeared to be loose, or appeared to be disconnecting from the gantry, do not use the system. Please contact your local BC Technical Field Service Engineer immediately. Enclosed is a verification form. Please fax the completed form at your earliest convenience to 415.663.4225. For further information or support concerning this issue, please call 925.922.2615.
- Initiated
- 2011-07-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2015-04-01
- Distribution
- Worldwide Distribution -- USA (nationwide) and the countries of France and United Kingdom
- Recall number
- Z-1214-2012
- Recall status
- Terminated
- CDRH record ID
- 107753
- FDA classified
- 2012-03-21
- Recall event number
- 61298