Skip to content

Public FDA record

Z-1214-2014: Product Brand Name(s): Alere Triage TOX+MTD Drug Screen, Catalog Number 94400, Lots W56013B and W56035B.

Recall record Z-1214-2014 for Alere San Diego, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1214-2014
Company in FDA record
Alere San Diego, Inc.
FDA classification date
Mar 17, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Alere San Diego, Inc. is initiating a recall of the Alere Triage TOX Drug Screen, Catalog Number 94400, Lots W56013B and W56035B because Alere's investigation has shown that these lots have a frequency of false negative results for PCP (Phencyclidine) that is higher than the frequency shown in the package insert. If you have questions about previously reported results.

Source facts

Action
Alere San Diego, Inc. sent out recall notification letters to customers via direct mail or email. The letter, dated February 3, 2014, titled "URGENT MEDICAL DEVICE RECALL", provided customers the following information: product description with codes, reason for recall, impact, contact information and customer/distributor required actions. A Customer/Distributor Verification Form was provided. Customers were instructed to discontinue use and discard the affected lots. Alternately, if you elect to continue to use the product customers must verify if it will not be used to report results for PCP. Customers were instructed to return the Customer Verification Form by fax to 858-805-8457 with the amount of product they have for credit. The information should be shared with their laboratory staff and the notification should be retained as part of their laboratory Quality System documentation. If the product has been forwarded customers were instructed to provide a copy of the letter to them. Customers with questions were instructed to contact Alere San Diego, Inc. at 877-308-8287 or email Triage.Support@alere.com. For questions regarding this recall call 858-805-2000, ext 3015.
Initiated
2014-02-05
Root cause
Process control
Terminated
2014-10-27
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Taiwan.
Recall number
Z-1214-2014
Recall status
Terminated
CDRH record ID
125676
FDA classified
2014-03-17
Recall event number
67529

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗